Heated
Static Headspace Injector
|
65
|
|
|
|
VOA Vial
Sleeve Thermostatted from Ambient to 90oC
Gas Sampling
Valve with fixed volume Loop
Gauge
displays Actual Vial Pressure
Liquid,
Solid, or Powder samples
Complete
PeakSimple Control
|
|
|
The
Heated Static Headspace injector is useful for the analysis of volatiles,
especially where the sample matrix is dirty. A 10-port
gas sampling valve and fixed volume sample loop are used for maximum
precision. |
|
The
thermostatted headspace sleeve accepts standard 40mL VOA vials containing
10-20mLs of sample.
|
|
Solenoid valve
|
As the vial
is inserted into the headspace sleeve, two needles puncture the
septum top of the vial. Purge gas enters through one needle to pressurize
the vial, and the other needle carries headspace vapors to the loop
of the gas sampling valve. A solenoid
valve located at the loop exit is opened under PeakSimple
data system control to allow headspace vapors to purge through
the loop just prior to injecting the loop contents onto the column.
The entire headspace sleeve is mechanically agitated under control
of the data system.
The headpsace
sleeve is thermostatted from ambient to 90oC
under PeakSimple control, and can be cooled down before removing
the VOA vial.
|
|
|
8690-0045
|
Heated
Static Headspace Injector
|
£5,448.00
|
|
|
Heated
Static Headspace Injector |
65
|
|
|
|