HT300A
110-Vial Liquid Autosampler
|
67
|
|
|
|
- Holds
110 2mL or 2.5mL vials
- Interfaces
with SRI and other GCs
- 15-Step
Automatic Injection Sequence
- Direct
Injection, No Transfer Lines
|
|
|
The HT300A Liquid
Autosampler is made to meet the high throughput liquid injection needs
of your GC analysis. Like the HT200H Headspace Autosampler, the swivel
head design simulates the movements of manual direct injection and
eliminates the need for transfer lines, as well as leaving the injection
port free for manual injections. Up to 10 analytical methods, including
function speeds, may be stored in the HT300A's memory. |
|
The
automatic injection sequence may have up to 15 steps, which may be
programmed to include: |
OPERATING
SPECIFICATIONS |
Sampling
Sample volume
Air volume
Aspirating speed
Needle washing
Air bubble removal
Viscosity time
Syringe sizes
Injection
Injection
speed
Waiting time before & after injection
Injection
depth
Internal Standard Technique
IS volume
Air gap volume
Mode
|
Sample
steps of 0.1uL
steps of 0.1uL
0.1uL - 100uL/second
up to 15
up to 15 strokes
1-60 seconds
1, 10, 25, 50, 100uL (1000uL large volume version)
Injection
0.1uL - 100uL/second
1-60 seconds
variable
IST
steps of 0.1uL
steps of 0.1uL
1 or 2 air gaps |
|
First sample
of group
Last sample of group
Analytical method
Number of injections for each sample
Pre and post washing solvent position
Internal standard position (if used)
|
The
sampling system eliminates air bubbles, and the variable fill speed
allows for a wide range of sample viscosities. The syringe may be
washed with solvent or sample. |
|
|
8690-3000
|
110-Vial
Liquid
Autosampler
|
Call
for quote: (0117) 230 5092
|
|
|
HT300A
110-Vial Liquid
Autosampler |
67
|
|
|
|