The Model 202 can
be mounted inside the 8610C, 8610D,
410, or 310 GCs and
the 210D HPLC, or it can be mounted in a separate
box, ready for connection to other manufacturers' GC or HPLC detectors.
The Model 202 data system consists of PeakSimple
chromatography software for Windows and a four channel, 24-bit high
resolution A/D board.
When mounted in an
SRI GC, the Model 202 controls the column oven
temperature program, and the pressure
program of the carrier gas electronic pressure controller (EPC). When
mounted in the Model 210D HPLC system, the
Model 202 controls the pump speed and gradient profile. The eight available
TTL outputs are connected internally within the GC or HPLC to control
functions such as valve rotation, gas solenoid actuation, autosampler
injection, etc.
When mounted in a
separate box, the temperature, pressure, and gradient control outputs
are available for use, but are not connected to anything. The eight TTL
outputs are wired to a bank of eight single-pole, dual-throw mechanical
relays with screw terminals for easy connection to any user device which
operates from a contact closure. Two remote start inputs allow run initiation
from the user's GC or HPLC system.
The four channels
of data can be randomly assigned to one of two time bases, which allows
independent start and stop times for two entirely separate instruments.
Data can be acquired at rates of up to 50Hz with one channel active, 10Hz
for two channels, or 5Hz with all four channels activated and acquiring
data. Windows computers with two available serial ports can operate dual
Model 202 systems, for a total of eight data channels and four time bases
from a single PC platform. For computers with USB ports only, we recommend
the IOGear USB PDA/Serial Adaptor (about $40US), which allows serial peripherals
to interface with USB computers.
|